Single-frequency CARS microscopy
Narrowband ps pump and Stokes pulses, tuned to one band, are raster-scanned. Video-rate lipid imaging is routine. Picosecond pulses (bandwidth ≈ linewidth) maximize the resonant-to-NR ratio.
A field guide for PhD students · nonlinear vibrational spectroscopy
Two laser colours beat at a molecular vibration, drive it coherently, and a third photon reads it out as a new, blue‑shifted beam. The signal is directional, bright and fluorescence‑free, and it comes with a coherent background you must understand to interpret any spectrum.
1 · The process
Solid lines are real molecular states; dashed lines are virtual states. Switch between CARS and its relatives to see how the same vibration is addressed differently.
The pump and Stokes fields beat at \(\omega_p-\omega_s\). When this beat matches a Raman-active vibration \(\Omega_R\), the molecules in the focal volume oscillate in phase. A probe photon (usually taken from the pump beam) scatters off this coherent vibration and leaves at \(\omega_{as}=2\omega_p-\omega_s\).
Because every molecule radiates in phase, the signal grows as \(N^2\) and leaves in a well-defined direction set by phase matching. It appears on the blue side of the pump, so one-photon fluorescence (which is red-shifted) is spectrally rejected.
The same four-wave mixing happens through purely electronic, instantaneous polarizability: no real vibrational level is involved. This gives a real, nearly frequency-independent \(\chi^{(3)}_{NR}\) at the same \(\omega_{as}\) and in the same direction.
It is not just an additive offset. It interferes with the resonant term and distorts the line. When \(2\omega_p\) or \(\omega_p+\omega_s\) approaches an electronic transition, two-photon enhancement makes the NRB much stronger, which is why near-IR excitation is preferred in microscopy.
The same two beams, but the signal is read at the input colours: the pump loses photons (stimulated Raman loss, SRL) and the Stokes gains them (SRG). The generated field is detected heterodyne against the input beam itself.
The measured modulation is \(\propto \mathrm{Im}\,\chi^{(3)}_R\,I_pI_s\). It is linear in concentration and has no non-resonant background to first order, though cross-phase modulation, transient absorption and thermal lensing can leak in.
A single pump photon scatters inelastically; the Stokes photon is emitted spontaneously into all directions with random phase. Intensities from different molecules add, so signal \(\propto N\).
Cross-sections are tiny, of order \(10^{-30}\,\mathrm{cm^2\,sr^{-1}}\) compared with \(\sim10^{-16}\,\mathrm{cm^2}\) for fluorescence. This is the core motivation for coherent Raman: drive the vibration instead of waiting for it.
2 · Theory
The classical Placzek model gives the right lineshape; the density-matrix treatment supplies the prefactor in terms of the Raman cross-section.
Expand the polarizability in the normal coordinate \(Q\): \(\alpha(Q)=\alpha_0+(\partial\alpha/\partial Q)_0\,Q+\dots\). The optical field exerts a force \(F=\tfrac12(\partial\alpha/\partial Q)_0E^2\), so
With \(E(t)=E_pe^{-i\omega_pt}+E_se^{-i\omega_st}+\mathrm{c.c.}\), the term oscillating at \(\Delta=\omega_p-\omega_s\) drives
The coherently vibrating molecules modulate the probe: \(P=N(\partial\alpha/\partial Q)_0\,Q\,E_p\). Collecting the \(\omega_{as}\) component (Boyd convention, degeneracy factor 3):
Third-order perturbation theory of the density matrix ties the amplitude to measurable quantities (Gaussian units, after Eckbreth):
where \(\Delta\rho\) is the fractional population difference between the lower and upper vibrational levels. This population dependence is exactly what makes CARS a thermometer (section 6).
Away from electronic resonances \(\chi_{NR}\) is real. The detected signal is the modulus squared:
Solving the slowly-varying-envelope wave equation for plane waves over length \(L\):
Three consequences dominate experiment design: quadratic in pump and linear in Stokes (so pulsed lasers with high peak power win); quadratic in \(N\) (so dilute species fade fast); and a phase-matching factor that sets geometry.
Mixing the anti-Stokes field with a phase-stable local oscillator makes the signal linear in \(\chi^{(3)}\). Choosing \(\varphi\) isolates \(\mathrm{Im}\,\chi^{(3)}\), which carries the spontaneous-Raman-like spectrum.
3 · Lineshape explorer
Model the CH₂ symmetric stretch of lipids at 2845 cm⁻¹. Raise \(\chi_{NR}\) and watch the peak move to lower wavenumbers while a dip opens on the high side. Lower the concentration and the dispersive cross-term takes over.
Units are arbitrary but consistent: each band has amplitude \(A=10N\) so that \(\mathrm{Im}\,\chi_R\) peaks at 1 for N = 1 and Γ = 10 cm⁻¹. In real biological samples \(\chi_{NR}\) from water and other solutes is often comparable to or larger than the resonant term for anything but lipids.
4 · Phase matching
Signal builds up only while the generated field stays in phase with the driving polarization, i.e. over about one coherence length.
Gases (spectroscopy, combustion). Dispersion is small but path lengths are millimetres to centimetres. Folded BOXCARS crosses the beams at small angles so the wave-vector diagram closes exactly. It also gives spatial resolution (the crossing volume) and spatially separates the signal.
Microscopy. A high-NA objective supplies a broad cone of wave vectors and the interaction length is only about 1 µm, much shorter than \(l_c\) in the forward direction. Collinear beams therefore work. The Gouy phase shift through focus adds a small, unavoidable mismatch.
Epi (backward) detection. Here \(\Delta k\approx 2k_{as}\), so \(l_c\approx\lambda_{as}/4n\). Bulk solvent signal cancels while objects smaller than about λ still radiate backwards. This is why epi-CARS highlights small features against an aqueous background.
Water dispersion from a two-term Cauchy fit, n(λ) = 1.3251 + 0.00247/λ²[µm], adequate for 0.6–1.3 µm. The y-axis is normalized to (2/Δk)².
5 · Time domain
The Fourier transform of the susceptibility is the impulse response. Each Raman mode is a damped oscillation, while the electronic NR response is effectively instantaneous:
A 10 cm⁻¹ line has \(T_2\approx1.06\) ps. If the probe arrives a few pulse widths after the pump–Stokes pair, the NR contribution, which follows the pulse cross-correlation, has gone. The vibrational coherence is still there.
Hybrid fs/ps CARS uses broadband fs pump and Stokes pulses to excite many modes at once, and a narrowband, time-delayed ps probe to read them out with spectral resolution. It has become a standard approach in gas-phase thermometry and in broadband microscopy.
The model plots \(|\chi_{NR}\,g(\tau)+R(\tau)|^2\), where \(g\) is a Gaussian cross-correlation and \(R\) is the exponential coherence convolved with \(g\). This is a heuristic envelope model that ignores chirp and quantum beats between modes.
6 · Gas-phase thermometry
The amplitude of each Q(J) line scales with the population difference \(\Delta\rho\). As the gas heats, high‑J lines (shifted to lower wavenumber by vibration–rotation coupling) and the v = 1→2 hot band grow. Fitting spectra like these is the workhorse of combustion diagnostics.
For ¹⁴N₂: \(\omega_e=2358.57\), \(\omega_ex_e=14.32\), \(B_e=1.998\), \(\alpha_e=0.0173\) cm⁻¹. Nuclear-spin weights are \(g_J=6\) for even J and 3 for odd J. The fundamental band head sits near 2330 cm⁻¹ and the hot band near 2301 cm⁻¹.
Simplified model: isotropic Q-branch only, constant Lorentzian width (0.5 cm⁻¹ HWHM, exaggerated for display), no Placzek–Teller or Herman–Wallis factors, no collisional narrowing. Real codes such as CARSFT or CARSpy include these.
7 · Bench toolbox
| Region | Shift (cm⁻¹) | Assignment | Typical use |
|---|---|---|---|
| Fingerprint | 1004 | Phenylalanine ring breathing | Protein marker, intensity reference |
| Fingerprint | 1080–1100 | PO₂⁻ sym. stretch | DNA/RNA backbone, nuclei |
| Fingerprint | 1440–1450 | CH₂ scissoring | Lipids and proteins |
| Fingerprint | 1650–1660 | Amide I / C=C stretch | Secondary structure, unsaturation |
| Silent | 2100–2250 | C–D, C≡C, C≡N | Isotope and alkyne tags |
| Gas | 2331 | N₂ Q-branch | Combustion thermometry |
| CH stretch | 2845–2850 | CH₂ symmetric stretch | Lipid droplets, myelin |
| CH stretch | 2930–2940 | CH₃ stretch | Proteins |
| High-wavenumber | 3200–3500 | O–H stretch | Water, hydration |
| Gas | 4155 | H₂ Q-branch | Fuel species, high-T thermometry |
8 · Implementations
Most variants exist to answer one of three problems: get a full spectrum, beat the non-resonant background, or recover the linear Raman response.
Narrowband ps pump and Stokes pulses, tuned to one band, are raster-scanned. Video-rate lipid imaging is routine. Picosecond pulses (bandwidth ≈ linewidth) maximize the resonant-to-NR ratio.
Narrowband pump plus a supercontinuum Stokes covers 500–3500 cm⁻¹ in one shot with a spectrometer. This yields hyperspectral images, but phase retrieval is needed to obtain Raman-like spectra.
Normalizing by a reference NRB and applying a Kramers–Kronig or maximum-entropy transform recovers \(\mathrm{Im}\,\chi^{(3)}\):
\(\varphi(\omega)=\mathcal{H}\{\ln\sqrt{I_{\text{CARS}}/I_{\text{NRB}}}\}\).
Backward detection with \(l_c\approx\lambda/4n\) suppresses bulk solvent and favours sub-wavelength objects. Turbid tissue also redirects forward signal back into the objective.
The NR tensor obeys Kleinman symmetry (depolarization ratio 1/3). Resonant bands generally do not, so an analyzer set orthogonal to the NR signal polarization rejects it at the cost of signal.
Rapidly switching \(\omega_p-\omega_s\) on and off resonance and lock-in detecting the difference removes the flat NRB and boosts sensitivity.
Delaying the probe beyond the pulse overlap removes the instantaneous NR response (section 5). This is standard in gas-phase thermometry.
Interference with a phase-locked local oscillator at \(\omega_{as}\) gives signal linear in \(\chi^{(3)}\) and in N. Choosing the LO phase separates \(\mathrm{Re}\) and \(\mathrm{Im}\).
One shaped fs pulse provides pump, Stokes and probe. Spectral-phase shaping selects which vibrational coherences interfere constructively.
ns or hybrid pulses with crossed beams provide point-wise temperature and major-species concentrations in flames, engines and plasmas, with mm-scale resolution.
9 · Side by side
| Property | Spontaneous Raman | CARS | SRS | CSRS |
|---|---|---|---|---|
| Detected frequency | \(\omega_p-\Omega_R\) | \(2\omega_p-\omega_s\) | \(\omega_p\) (loss) or \(\omega_s\) (gain) | \(2\omega_s-\omega_p\) |
| Measures | \(\propto\mathrm{Im}\,\chi_R\) | \(|\chi_{NR}+\chi_R|^2\) | \(\mathrm{Im}\,\chi_R\) | \(|\chi_{NR}+\chi_R^{*}|^2\) |
| Concentration scaling | \(N\) | \(N^2\) (\(N\) when NR dominates) | \(N\) | \(N^2\) |
| Power scaling | \(I_p\) | \(I_p^2I_s\) | \(I_pI_s\) | \(I_pI_s^2\) |
| Non-resonant background | None (fluorescence is the issue) | Yes, interferes | Largely absent | Yes |
| Lineshape | Lorentzian | Dispersive, red-shifted peak | Lorentzian, matches Raman | Dispersive, mirror of CARS |
| Fluorescence immunity | Poor | Good (blue-shifted signal) | Good (modulation transfer) | Poor (red side) |
| Detection | Spectrometer + CCD | PMT / spectrometer, filter at new colour | Photodiode + lock-in (MHz modulation) | Spectrometer |
| Imaging speed | ms–s per pixel | µs per pixel | µs per pixel | Rarely used for imaging |
10 · Self-check
Try answering before opening each one. Several can be checked with the models above.
\(\mathrm{Re}\,\chi_R\propto(\Omega_R-\Delta)\) is positive below resonance and negative above. The cross-term \(2\chi_{NR}\mathrm{Re}\,\chi_R\) therefore adds below and subtracts above. At exact resonance \(\mathrm{Re}\,\chi_R=0\), so the cross-term vanishes there. Check it in the lineshape explorer.
\(I_{as}\propto I_p^2\) gives ×4 from the pump, and \(|\chi_R|^2\propto N^2\) gives ×4 from concentration, so ×16 in the resonance-dominated limit. When the NRB dominates, the useful contrast is the cross-term, which is linear in N, so the concentration factor is only ×2.
SRS is self-heterodyned with the input beam at the same frequency. The detected modulation is \(\propto\mathrm{Im}\,\chi^{(3)}\). A real, instantaneous \(\chi_{NR}\) produces no net energy exchange between pump and Stokes, so it does not appear. Residual artefacts come from cross-phase modulation, transient absorption and photothermal effects.
Backward \(\Delta k\approx 2k_{as}\) gives \(l_c\approx\lambda_{as}/4n\approx 0.12\) µm. Contributions from a thick, uniform medium cancel by destructive interference. An object comparable to or smaller than \(l_c\)–λ radiates without full cancellation.
The NR response is driven by the whole pulse bandwidth, but the resonance uses only about 10 cm⁻¹ of it. The resonant/NR ratio collapses and spectral resolution is lost. Fixes: use ps pulses matched to the linewidth, spectral focusing (equally chirped pump and Stokes), or a delayed narrowband probe (hybrid fs/ps).
Line amplitudes follow the population difference \(\Delta\rho_{v,J}\). Rotational redistribution to high J broadens the Q-branch towards lower wavenumber, and the v = 1→2 hot band appears near 2301 cm⁻¹ above roughly 1000 K. A least-squares fit of a theoretical library to the measured spectrum returns T, typically to a few percent.
11 · Further reading